LS125 UV Light Meter+UVA-X2 Probe
Suitable for 340nm UVA aging lamps
340nm aging lamps intensity and energy detection
Large range of up to 200,000 μW/cm²
Multiple unit selection and intelligent statistics
The LS125 UV radiometer host is an intelligent host that supports 12 different probes, primarily for data display. The UVA-X2 probe has a spectral response range of 315nm-365nm, specializing in detecting the light intensity and energy of 340nm UVA aging lamps, mainly in UV aging test chambers.
If you want to know about other types of probes supported by the LS125 UV light meter host, please click "LS125 Multi-Probe UV light meter Host" for details.
Parameters
Parameters | UVA-X2 Probe |
---|---|
Spectral Response | 315nm-365nm,λp = 340nm |
Power Measuring Range | 0 - 200000μW/cm² |
Resolution | 0.1μW/cm² |
Energy Measuring Range | 0 - 9999999μJ/cm² |
Record Time | 0 - 99999S |
Measuring Accuracy | H<50μW/cm²:±5μW/cm², H>=50μW/cm²:±10%H,(H is the standard value) |
Sampling Speed | 6times/second |
Optional Unit | μW/cm² (default),mW/cm²,W/m² |
Test Hole Diameter | Φ10mm |
Probe Size | 39mm × 32mm × 15mm (L*W*H) |
Features
Specialized for 340nm UVA aging lamps detection
Spectral response range is 315nm-365nm, with a peak wavelength of 340nm, designed for xenon arc lamp light detection.
Large range of up to 200,000 μW/cm²
Maximum measuring power reaches 200,000 μW/cm².
Advanced digital probe, anti-interference
It adopts digital probe to collect and process the data which makes it anti-interference and more accurate measurement.
Intelligent data statistics, worry-free
Rich statistical functions display real-time power values, maximum, minimum, average values, time, and energy values simultaneously, no need to manually record and calculate.
Switch units at will, more convenient
Three units μW/cm², mW/cm², W/m² can be switched as needed, with data conversion between different units without manual calculation.
Qualifications and Standards
The UV light meter LS125 host + UVC-X2 probe conform to the following standard:
JJG 879-2015 Verification Regulation of Ultraviolet Radiometers
FAQ
-
1. The difference between UVA-X0, UVA-X1 and UVA-X2 UV probe for LS125 UV light meter host
1) UVA-X0 probe is mainly used to detect the radiation intensity and energy of 365nm high pressure mercury lamp, the spectral response range is 315-400nm, the power measuring range is 0-2000mW/cm²;
2) The UVA-X1 probe is mainly used to detect the radiation intensity and energy of low-power UVA light sources, the spectral response range is 260-400nm, the power measuring range is 0-200000μW/cm²;
3) The UVA-X2 probe is mainly used to detect the radiation intensity and energy of 340nm UVA aging lamps, the spectral response range is 315nm-365nm, the power measuring range is 0-200000μW/cm².
-
2. What's the working temperature of the UV probes?
The 12 probes of the LS125 UV light meter are not heat-resistant. The probe can be used for a long time at 80℃ or lower. If the temperature is higher than 80℃, it is recommended to use UV energy meter for UVA light source detection.
-
3. Which probe is recommended if I need to measure both mercury lamp and LED light source?
It is recommended to use the LS125 UV light meter host and UVA-X0, UVALED-X0 probes.
-
4. Whether a traceable calibration certificate is included?
UVA-X0 and UVA-X1 are mainly used to detect 365nm high-pressure mercury lamps, UVA-X2 is mainly used to detect the radiation intensity and energy of 340nm UVA aging lamps. They both ensured to pass the inspection of the Metrology Institute and obtain a calibration certificate (approved by CNAS and ilac-MRA). The cost of the calibration certificate is about US$200;
But the UVALED-X0, UVALED-X1 and UVALED-X3 probes are professionally used to detect UVA LED light sources. There is no unified inspection standard for LED light sources, each light source and UV energy meter manufacturer has their own calibration standards. We can provide the factory calibration reports for these three probes special for UV LED, but no calibration certificates can be provided.
Your product information request
Please let us know your requirements and we will be in contact with you shortly!